The role of percutaneous interventions in complex coronary disease and complex patients has been steadily expanding over the recent years. Left main and multi-vessel coronary artery interventions post a unique set of challenges to interventional cardiologists; especially in elderly patients, patients with left ventricular dysfunction or renal insufficiency. This resulted in an ongoing need for a safe and effective circulatory device that is easy to insert and manage in order to reduce the risk of morbidity and mortality associated with these interventions. Here, we describe a case of multi-vessel intervention including the left main coronary artery in a patient with reduced ejection fraction and multiple occluded bypass grafts.
A Ahmed, Use of Impella 2.5 device for … http://dx.doi.org/10.25239/SHJ/Vol5/No1/casereport2
Use of Impella 2.5 device for hemodynamic support during a High Risk Unprotected Left main percutaneous coronary intervention Ahmed S Ahmed, M.D. Interventional Cardiologist St Bernard Heart and Vascular, Jonesboro, AR, USA. Email: drahmedkagam@yahoo.com
The role of percutaneous interventions in complex coronary disease and complex patients has been steadily expanding over the recent years. Left main and multi-vessel coronary artery interventions post a unique set of challenges to interventional cardiologists; especially in elderly patients, patients with left ventricular dysfunction or renal insufficiency. This resulted in an ongoing need for a safe and effective circulatory device that is easy to insert and manage in order to reduce the risk of morbidity and mortality associated with these interventions. Here, we describe a case of multi-vessel intervention including the left main coronary artery in a patient with reduced ejection fraction and multiple occluded bypass grafts.
Keywords: Impella –hemodynamic support –high risk PCI – assist device
Case History:
A 69 years old male with hypertension, hyperlipidemia and coronary artery disease (CAD) with history of coronary artery bypass grafting (CABG) with left internal mammary artery (LIMA) to the left anterior descending artery (LAD) and reverse saphenous vein grafts (SVG) to the first diagonal artery and the first obtuse marginal (OM) artery presents with unstable angina 18 months after CABG. He was on 3 antiangina agents and optimal medical therapy for CAD, there were no ECG changes and troponin levels were negative. Given his clinical presentation and past history, cardiac catheterization was performed. It showed 60-70% stenosis in the distal left main coronary artery (LMCA) (Photo 1), 50% ostial and 70% mid LAD stenoses, 90% stenosis in the proximal first diagonal, 80% ostial left circumflex artery (LCX) stenosis and a 70% stenosis in the second OM branch. The right coronary artery had mild to moderate disease. All bypass grafts were occluded except for SVG to the first diagonal. It was noticed that the native coronary arteries were small in caliber based on angiographic appearance and previous description in the CABG operative report. Left ventricular ejection fraction was moderately reduced at 40% with moderate global hypokinesis. Myocardial perfusion imaging showed ischemia along the LCX and distal LAD territory. The decision was made to proceed with multi-vessel percutaneous coronary intervention (PCI) with Impella® (Abiomed Inc, Danvers, MA, USA) hemodynamic support.
A Ahmed, Use of Impella 2.5 device for … http://dx.doi.org/10.25239/SHJ/Vol5/No1/casereport2
Photo 1
Photo 2
Right common femoral access was obtained Impella device was inserted via left common and a pigtail catheter was used to obtain an femoral access using a 13 French sheath and abdominal aortogram with bilateral iliac placement was confirmed both with angiograms. The size of the left iliac artery fluoroscopy and console placement signal was adequate for Impella® 2.5 insertion. The A Ahmed, Use of Impella 2.5 device for … http://dx.doi.org/10.25239/SHJ/Vol5/No1/casereport2 (Photo2). The device provided blood flow between 1.8 and 2.0 liters per minute. This flow rate was adequate for hemodynamic support for the patient’s body surface area (1.77 m2).
The LMCA was engaged with a 6 Fr Judkins Left 4.0 guiding catheter and the mid LAD LMCA lesion was then pre-dilated with a 2.0x15 mm compliant balloon utilizing the wire into the LCX. There was a mild drop in blood pressure with balloon inflation but no change in Impella flow. A 2.5x20 mm drug lesion was crossed with a Marvel®guide eluting stent was deployed in the LMCA into the LCX. There was 70% residual wire (Boston Scientific Corp – ostial LAD stenosis after stent deployment. Marlborough, MA, USA). The lesion was The LAD wire was then retracted and used pre-dilated with a 2.0x15 mm compliant to re-wire the LAD through the LMCA stent balloon and then a 2.25x32 mm struts. The LMCA stent was post-dilated Synergy®drug eluting stent (Boston with a 2.75x15 mm non-compliant balloon. Scientific Corp – Marlborough, MA, USA) The ostial LAD was dilated with a 2.5x12 was deployed. Post dilation was performed compliant balloon with 30% residual with a 2.5x20mm non-compliant balloon.
stenosis (Photo3).
The wire was left in place in the LAD and a The patient was transferred to the intensive second guide wire was used to cross the care unit and the Impella catheter and sheath mid-LCX lesion. This lesion was prewere removed with good haemostasis using dilated with a 2.0x15 mm compliant balloon manual pressure. He was discharged the then a 2.5x24 mm Synergy®drug eluting next day in good condition. At one week and six-week outpatient follow-up visits, he had stent was deployed and post-dilated with a no residual angina.
2.5x20 mm non-compliant balloon. The
Photo 3 A Ahmed, Use of Impella 2.5 device for … http://dx.doi.org/10.25239/SHJ/Vol5/No1/casereport2
surgical cut-down (these pumps provide up Complete coronary revascularization in to 2.5 liters, 3.5 liters and 5 liters of flow per patients with multi-vessel CAD was minute, respectively). The pump is powered associated with lower mortality, myocardial and controlled by an external console(7). infarction and repeat revascularization The pump motor diameter is 12 French (Fr) compared to incomplete revascularization for Impella 2.5, 14 Fr for Impella CP and 21 (1, 2). Although CABG remains the Fr for Impella 5.0(8). We used Impella 2.5 preferable revascularization strategy in such device in our patient given the small size of patients the role of percutaneous coronary the left iliac artery.
intervention (PCI) has been expanding in the The safety of Impella 2.5 device use for high recent years. This is especially true in risk PCI was demonstrated by several patients with severe co-morbidities, renal studies showing no limb ischemia, valve insufficiency, advanced age, reoperation or trauma or aortic regurgitation (9, 10) and poor distal targets which make surgery a less only a small incidence of device related attractive option(3). PCI in this patient bleeding.
population has several challenges both from In PROTECT-II trial, it was concluded that a technical and logistical standpoint with hemodynamic support with Impella longer procedure times, larger contrast compared with intra-aortic balloon pump volumes and hemodynamic compromise.
during high risk PCI resulted in improved The elective use of Intra-Aortic Balloon event free survival at 3 months follow- Pump (IABP), as the usual hemodynamic up(12).
support device, in high risk PCI has not
shown a significant and reproducible The use of Impella for hemodynamic reduction in adverse cardiac outcomes in support in high risk PCI is well described in several studies (4,5).
the literature. This case report illustrates that The ideal cardiac support device for the PCI remains a valid option for patients who cardiac catheterization laboratory provides are at high risk after early multi-graft failure. effective support for both elective high-risk This patient’s option would have otherwise PCI, as well as cardiogenic shock. For been unprotected high risk intervention or elective high-risk patients, the goal is to redo CABG with high potential graft failure. maintain stable hemodynamics, provide The Use of intravascular ultrasound would more time for complex PCI by raising the have been a reasonable step to take for patient’s ischemic threshold to minimize accurate sizing of the LMCA.
myocardial cell damage from balloon inflation or coronary dissection; and provide prophylactic safety in case of complications